Talk:Comparison of the Measuring Method: Difference between revisions
SLeithaeuser (talk | contribs) No edit summary |
SLeithaeuser (talk | contribs) No edit summary |
||
Line 24: | Line 24: | ||
|bgcolor = "#FFFFEO" | 0 - 40 °C | |bgcolor = "#FFFFEO" | 0 - 40 °C | ||
|- | |- | ||
|bgcolor = "#F7F7F7" | ''' | |bgcolor = "#F7F7F7" | '''Capacitive <br>(Al-Oxide)''' | ||
|bgcolor = "#FFFFEO" | change in capacitance | |bgcolor = "#FFFFEO" | change in capacitance | ||
|bgcolor = "#FFFFEO" | a | |bgcolor = "#FFFFEO" | a | ||
Line 33: | Line 33: | ||
|- | |- | ||
|bgcolor = "#F7F7F7" | '''LiCl''' | |bgcolor = "#F7F7F7" | '''LiCl''' | ||
|bgcolor = "#FFFFEO" | LiCl | |bgcolor = "#FFFFEO" | LiCl equilibrium temperature | ||
|bgcolor = "#FFFFEO" | a | |bgcolor = "#FFFFEO" | a | ||
|bgcolor = "#FFFFEO" | g | |bgcolor = "#FFFFEO" | g | ||
Line 40: | Line 40: | ||
|bgcolor = "#FFFFEO" | -10 - 50 °C | |bgcolor = "#FFFFEO" | -10 - 50 °C | ||
|- | |- | ||
|bgcolor = "#F7F7F7"| ''' | |bgcolor = "#F7F7F7"| '''Infrared''' | ||
|bgcolor = "#FFFFEO" | absorption | |bgcolor = "#FFFFEO" | absorption of infrared light | ||
|bgcolor = "#FFFFEO" | a | |bgcolor = "#FFFFEO" | a | ||
|bgcolor = "#FFFFEO" | g,l,s | |bgcolor = "#FFFFEO" | g,l,s |
Revision as of 10:47, 5 November 2012
Author: Hans-Jürgen Schwarz
back to Air Humidity Measurement
The table lists the different measuring principles, evaluates them and shows the different criteria for their application.
Name | Principle | Properties | Measuring range | Temperature range | ||
Chilled mirror hygrometer | Condensation on a chilled mirror | a | g | + very accurate - expensive; prone to damage by pollution |
-40 - +50 °C dew point |
0 - 40 °C |
Capacitive (Al-Oxide) |
change in capacitance | a | g,l | + small, low cost, fast - drift, chemically sensitive, calibration |
-100 - +60 °C dew point |
0 - 70 °C |
LiCl | LiCl equilibrium temperature | a | g | + robust, grease resistant - continuous operation required |
12 - 100% RH | -10 - 50 °C |
Infrared | absorption of infrared light | a | g,l,s | + solid, contact free - installation, expensive |
0 - 30% water content |
- |
Microwaves | absorption of microwaves | a | g,l,s | + solids, contact free - installation, expensive |
1 - 1000 g/m² | - |
Conductivity | measuring the electrical conductivity | a | s | + solids (wood, concrete), hand held - accuracy |
6 - 100% water content |
- |
Psychrometer | temperature difference wet/dry | r | g | + robust - needs maintenance |
0 - 100% RH | 0 - 50 °C |
Hygrometer | change in length of hair or synthetic fiber | r | g | + direct display, risible - inaccurate, sensitive to dust |
0 - 100% RH | -30 - 110 °C |
Capacitive | polymers change in capacitance | r | g | +low price, small, quick - Drift, chemically sensitive, calibration |
0 - 100% RH | -40 - 180 °C |
Electrolytic | current flow through phosphorous oxide | a | g | + no calibration, ppm range - flow rate, measuring range |
5 - 1000 ppm | 0 - 40 °C |
Resistive | Resistance of a hygroscopic layer | r | g | + low cost, good measuring effect, hysteresis - condensation, temperature influence |
30 - 90% RH | 10 - 40 °C |
a - absolute, r - relative, g - in gas, l - in liquids, s - in solids
Time response[edit]
Criteria for the usefulness of sensors are:
- the accuracy of the measurement,
- the transition function,
but also the
- time response.
The time response for most sensors is mainly determined by the amount of water needed for equilibrium. However, the time the water takes to diffuse onto the sensor needs to considered, too. Since this amount of water can be provided or discharged by the ambient air, the amount of air that flows onto a sensor plays an important role.
Because the diffusion rate is a function of temperature, a sensor will always have a slower reaction the lower the temperature. Other factors apply for sensors with different measuring principles, e.g. chilled mirror hygrometer.
In addition, filters complicate the exchange of gases. When using filter sizes of below 5 - 8 µm no more direct gas exchange takes place, but only gas diffusion.
Type | T50-time | Interval skip??????? |
resistive | < 60 s | 30 / 80 % RH |
capacitive | < 5 min | 30 / 80 % R |
LiCl | < 5 min | -1 / 10 °C TP |
Chilled mirror hygrometer | < 60 s | -20 / 0 °C TP |
elektrolytic | < 30 min | 0 / 100 ppm |
Fiber | < 10 min | 10 / 90 % RH |
SLeithaeuser 11:43, 5 November 2012 (CET)