Micro-chemical test for magnesium

From Saltwiki
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.

Author: Christine Bläuer

back to Micro-chemical testing

For more information on the required equipment and materials see the Micro-chemistry article.

Procedure

Mg2+ test using ammonium hydrogen phosphate and ammonia

Acidify the sample drop with a small drop of dilute (2M) hydrochloric acid. Add a drop of di-ammonia hydrogen phosphate solution (10% aqueous solution). Invert the slide and hold the drop over an open bottle of concentrated ammonium hydroxide so that the ammonia vapor reaches it. Alternatively a drop of 2M ammonium hydroxide solution can be added to the test drop, but the first procedure usually gives better results.

If Mg2+ is present typical crystals of magnesium ammonium phosphate (MgNH4PO4.6H2O) can be observed in the center of the test drop. Detection limit for Mg2+ = 0.05 µg.

Ions that may inhibit the test or affect its sensitivity: If calcium ions are present, these form a white, flocculent precipitate of calcium phosphate which can interfere or even inhibit the formation of magnesium ammonium phosphate

Mg2+ test using cesium chloride and sodium hydrogen phosphate

Acidify the sample drop with a small drop of dilute (2M) hydrochloric acid. Add to it a drop of di-sodium hydrogen phosphate solution (10% aqueous solution). Add one crystal of cesium chloride to the solution.

If Mg2+ is present, very small octahedral crystals of cesium magnesium phosphate (CsMgPO4.6H2O) will quickly form near the dissolving grain of CsCl. Detection limit for Mg2+ = 1 µg.

Ions that may inhibit the test or affect its sentivity: Ions of the elements Sr, Ba, Zn, Cd, Sn, Pb, Ag and Cu interfere with or mask the reaction. Calcium ions form a white, flocculent precipitate of calcium phosphate that interferes or even inhibits the formation of the cesium magnesium phosphate formed.

Mg2+ test using Titan yellow, a triazin dye, after [1]

Put 1 drop of the test solution on a spot plate, acidify with diluted (2M) hydrochloric acid and 1 drop of the titan yellow solution. Add 1 or 2 drops of sodium hydroxide solution and stir with a glass rod. If Mg2+ is present a red, flocculent precipitate will form. Detection limit for Mg2+ ca. 1.5 µg

Ions that may inhibit the test or affect its sentivity: The presence of calcium leads to a slightly orange to reddish staining of the solution which cannot be confused with the red, flocculent precipitation. Control tests with known substances will help to recognize a positive reaction for Mg2+ unambiguously.

Nickel, zinc, manganese and cobalt ions interfere with the reaction [2]

Literature